In vitro evaluation of the method effectiveness to limit inflation pressure cuffs of endotracheal tubes

Avaliação in vitro da eficácia de método para limitar a pressão de insuflação dos balonetes das cânulas endotraqueais

Rafael de Macedo Coelho; Thiago Trigueiro Morais de Paiva; Ligia Andrade da Silva Telles Mathias

Downloads: 0
Views: 899

Abstract

ABSTRACT BACKGROUND AND OBJECTIVE: Cuffs of tracheal tubes protect the lower airway from aspiration of gastric contents and facilitate ventilation, but may cause many complications, especially when the cuff pressure exceeds 30 cm H2O. This occurs in over 30% of conventional insufflations, so it is recommended to limit this pressure. In this study we evaluated the in vitro effectiveness of a method of limiting the cuff pressure to a range between 20 and 30 cm H2O. METHOD: Using an adapter to connect the tested tube to the anesthesia machine, the relief valve was regulated to 30 cm H2O, inflating the cuff by operating the rapid flow of oxygen button. There were 33 trials for each tube of three manufacturers, of five sizes (6.5-8.5), using three times inflation (10, 15 and 20 s), totaling 1485 tests. After inflation, the pressure obtained was measured with a manometer. Pressure >30 cm H2O or <20 cm H2O were considered failures. RESULTS: There were eight failures (0.5%, 95% CI: 0.1-0.9%), with all by pressures <20 cm H2O and after 10 s inflation (1.6%, 95% CI: 0 5-2.7%). One failure occurred with a 6.5 tube (0.3%, 95% CI: -0.3 to 0.9%), six with 7.0 tubes (2%, 95% CI: 0.4-3.6%), and one with a 7.5 tube (0.3%, 95% CI: -0.3 to 0.9%). CONCLUSION: This method was effective for inflating tracheal tube cuffs of different sizes and manufacturers, limiting its pressure to a range between 20 and 30 cm H2O, with a success rate of 99.5% (95% CI: 99.1-99.9%).

Keywords

Method, Inflation, Cuffs, Tube, Pressure

Resumo

RESUMO JUSTIFICATIVA E OBJETIVO: Os balonetes das cânulas traqueais protegem as vias aéreas inferiores da aspiração de conteúdo gástrico e facilitam a ventilação pulmonar, mas podem provocar diversas complicações, principalmente quando a pressão do balonete supera 30 cm H2O. Isto ocorre em mais de 30% das insuflações convencionais, sendo recomendada a limitação desta pressão. Neste estudo avaliou-se in vitro a eficácia de um método para limitar a pressão dos balonetes à faixa entre 20 e 30 cm H2O. MÉTODO: Utilizando um adaptador para conectar a cânula testada ao aparelho de anestesia, regulou-se a válvula limitadora deste a 30 cm H2O, insuflando o balonete por meio do acionamento do botão de fluxo rápido de oxigênio. Realizaram-se 33 testes para cada cânula de três fabricantes, de cinco tamanhos (6.5 a 8.5), utilizando três tempos para insuflação (10, 15 e 20 segundos), totalizando 1485 testes. Terminada a insuflação, mediu-se a pressão obtida com um manômetro. Pressões >30 cm H2O ou < 20 cm H2O foram consideradas falhas. RESULTADOS: Ocorreram oito falhas (0,5%; IC 95%: 0,1-0,9%), sendo todas por pressões <20 cm H2O e após insuflações de 10 segundos (1,6%; IC 95%: 0,5-2,7%). Uma falha ocorreu com cânula 6.5 (0,3%; IC 95%: -0,3-0,9%), seis com cânulas 7.0 (2%; IC 95%: 0,4-3,6%), e uma com cânula 7.5 (0,3%; IC 95%: -0,3-0,9%). CONCLUSÃO: Este método mostrou-se eficaz para insuflar os balonetes de cânulas traqueais de diferentes tamanhos e fabricantes limitando sua pressão à faixa entre 20 e 30 cm H2O, com incidência de sucesso de 99,5% (IC 95%: 99,1-99,9%).

Palavras-chave

Método, Insuflação, Balonetes, Cânulas, Pressão

References

Estebe JP, Delahaye S, Le Corre P. Alkalinization of intracuff lidocaine and use of gel lubrication protect against tracheal tube-induced emergence phenomena. Br J Anaesth. 2004;92:361-6.

Kaneda N, Goto R, Ishijima S. Laryngeal granuloma caused by short-term endotracheal intubation. Anesthesiology. 1999;90:1482-3.

Audu P, Artz G, Scheid S. Recurrent laryngeal nerve palsy after anterior cervical spine surgery. Anesthesiology. 2006;105:898-901.

Liu J, Zhang X, Gong W. Correlations between controlled endotracheal tube cuff pressure and postprocedural complications a multicenter study. Anesth Analg. 2010;111:1133-7.

Blanch PB. Laboratory evaluation of 4 brands of endotracheal tube cuff inflator. Respir Care. 2004;49:166-73.

Crimlisk JT, Horn MH, Wilson DJ. Artificial airways a survey of cuff management practices. Heart Lung. 1996;25:225-35.

Galinski M, Tréoux V, Garrigue B. Intracuff pressures of endotracheal tubes in the management of airway emergencies the need for pressure monitoring. Ann Emerg Med. 2006;47:545-7.

Sengupta P, Sessler DI, Maglinger P. Endotracheal tube cuff pressure in three hospitals, and the volume required to produce an appropriate cuff pressure. BMC Anesthesiol. 2004;4:8-13.

Tobias JD, Schwartz L, Rice J. Cuffed endotracheal tubes in infants and children should we routinely measure the cuff pressure?. Int J Pediatr Otorhinolaryngol. 2012;76:61-3.

Dobrin PB, Goldberg EM, Canfield TR. The endotracheal cuff a comparative study. Anesth Analg. 1974;53:456-60.

Chapman J, Pallin D, Ferrara L. Endotracheal tube cuff pressures in patients intubated before transport. Am J Emerg Med. 2009;27:980-2.

Felten ML, Schmautz E, Delaporte-Cerceau S. Endotracheal tube cuff pressure is unpredictable in children. Anesth Analg. 2003;97:1612-6.

Bassi M, Zuercher M, Erne JJ,. Endotracheal tube intracuff pressure during helicopter transport. Ann Emerg Med. 2010;56:89-93.

Yildirim ZB, Uzunkoy A, Cigdem A. Changes in cuff pressure of endotracheal tube during laparoscopic and open abdominal surgery. Surg Endosc. 2012;26:398-401.

Stanley TH, Foote JL, Liu WS. A simple pressure relief valve to prevent increases in endotracheal tube cuff pressure and volume in intubated patients. Anesthesiology. 1975;43:478-81.

Slocum Jr AH, Slocum Sr AH, Spiegel JE. Technical communication design and in vitro testing of a pressure-sensing syringe for endotracheal tube cuffs. Anesth Analg. 2012;114:967-71.

Kim JM. The tracheal tube cuff pressure stabilizer and its clinical evaluation. Anesth Analg. 1980;59:291-6.

Gravenstein N, Burwick N. Recoil of inflation syringe plunger limits excessive endotracheal tube cuff pressure. Anesthesiology. 1988;69:A730.

Doyle DJ. Digital display of endotracheal tube cuff pressures made simple. Anesthesiology. 1999;91:329.

Resnikoff E, Katz JA. A modified epidural syringe as an endotracheal tube cuff pressure-controlling device. Anesth Analg. 1990;70:208-11.

Lenoir RJ. Venous congestion of the neck; its relation to laryngeal mask cuff pressures. Br J Anaesth. 2004;93:476-7.

Miller RA. A new laryngoscope. Anesthesiology. 1941;2:317-20.

Spiegel JE. Endotracheal tube cuffs design and function. Anesthesiol News. 2010;36:51-8.

Nordin U. The trachea and cuff-induced tracheal injury an experimental study on causative factors and prevention. Acta Otolaryngol Suppl. 1977;345:1-71.

Combes J, Schauvliege F, Peyrouset O. Intracuff pressure and tracheal morbidity influence of filling cuff with saline during nitrus oxide anesthesia. Anesthesiology. 2001;95:1120-4.

Lewis FR, Schlobohm RM, Thomas AN. Prevention of complications from prolonged tracheal intubation. Am J Surg. 1978;35:452-7.

Lomholt N. Tracheal models which simulate the compliance of the adult male trachea. Br J Anaesth. 1992;69:98-100.

5dcd8de40e88258c11bf58f2 rba Articles
Links & Downloads

Braz J Anesthesiol

Share this page
Page Sections