Pacemakers and implantable cardioverter defibrillators - general and anesthetic considerations
Marcapassos e cardioversores desfibriladores implantáveis - considerações gerais e anestésicas
Amy G. Rapsang; Prithwis Bhattacharyya
Abstract
Keywords
Resumo
Palavras-chave
Referencias
Rozner MA. Implantable cardiac pulse generators: pacemakers and cardioverter-defibrillators. Miller's anesthesia. :1402-1388.
Bernstein AD, Daubert JC, Fletcher RD. The revised NASPE/BPEG generic code for antibradycardia, adaptive-rate, and multisite pacing. North American Society of Pacing and Electrophysiology/British Pacing and Electrophysiology Group.. Pacing Clin Electrophysiol. ;25:260-264.
ACC/AHA/HRS 2008 guidelines for device-based therapy of cardiac rhythm abnormalities. J Am Coll Cardiol. ;51:1-62.
Banker R, Mitchell R, Badhwar N. Pacemakers and implantable cardioverter-defibrillator emergencies. Cardiac intensive care. :334-310.
Sethuran S, Toff WD, Vuylsteke A. Implanted cardiac pacemakers and defibrillators in anaesthetic practice. Br J Anaesth. ;88:631-627.
Practice Advisory for the Perioperative Management of Patients with Cardiac Rhythm Management Devices: Pacemakers and Implantable Cardioverter-Defibrillators. A Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Rhythm Management Devices.. Anesthesiology. ;103:186-198.
Kleinman B, Hamilton J, Hariman R. Apparent failure of a precordial magnet and pacemaker programmer to convert a DDD pacemaker to VOO mode during the use of the electrosurgical unit. Anesthesiology. ;86:247-250.
Rastogi S, Goel S, Tempe DK. Anaesthetic management of patients with cardiac pacemakers and defibrillators for noncardiac surgery. Ann Cardiac Anaesth. ;8:32-21.
Chien WW, Foster E, Phillips B. Pacemaker syndrome in a patient with DDD pacemaker for long QT syndrome. Pacing Clin Electrophysiol. ;14:1212-1209.
Brode SE, Schwartzman D, Callans DJ. ICD-antiarrhythmic drug and ICD-pacemaker interactions. J Cardiovasc Electrophysiol. ;8:830-842.
Kalin R, Stanton MS. Current clinical issues for MRI scanning of pacemaker and defibrillator patients. PACE. ;28:328-326.
Levine GN, Gomes AS, Arai AE. Safety of magnetic resonance imaging in patients with cardiovascular devices: an American Heart Association scientific statement from the Committee on Diagnostic and Interventional Cardiac Catheterization. Circulation. ;116:2878-2891.
Roguin A, Goldsher D. Magnetic resonance imaging in individuals with cardiovascular implantable electronic devices. Europace. ;10:346-336.
Schwartzenburg CF, Wass CT, Strickland RA. Rate-adaptive cardiac pacing: implications of environmental noise during craniotomy. Anesthesiology. ;87:1254-1252.
Hayes DL, Holmes DR, Gray JE. Effect of 1.5 Tesla nuclear magnetic resonance imaging scanner on implanted permanent pacemakers.. J Am Coll Cardiol. ;10:786-782.
Sommer T, Vahlhaus C, Lauck G. MR imaging and cardiac pacemakers: in-vitro evaluation and in-vivo studies in 51 patients at 0.5 T.. Radiology. ;215:869-879.
Gimbel JR, Johnson D, Levine PA. Safe performance of magnetic resonance imaging on five patients with permanent cardiac pacemakers. Pacing Clin Electrophysiol. ;19:919-913.
Nazarian S, Roguin A, Zviman MM. Clinical utility and safety of a protocol for noncardiac and cardiac magnetic resonance imaging of patients with permanent pacemakers and implantable-cardioverter defibrillators at 1.5 Tesla.. Circulation. ;114:1284-1277.
Kanal E, Barkovich AJ, Bell C. ACR guidance document for safe MR practices. Am J Roentgenol. ;188:1-27.
Martin TE, Coman JA, Shellock FG. Magnetic resonance imaging and cardiac pacemaker safety at 1.5- Tesla.. J Am Coll Cardiol. ;43:1315-1324.
Loewy J, Loewy A, Kendall EJ. Reconsideration of pacemakers and MR imaging. Radiographics. ;24:1257-1268.
Domino KB, Smith TC. Electrocautery-induced reprogramming of a pacemaker using a precordial magnet. Anesth Analg. ;62:612-609.
Belott PH, Sands S, Warren J. Resetting of DDD pacemakers due to EMI. Pacing Clin Electrophysiol. ;7:169-172.
Van Hemel NM, Hamerlijnck RP, Pronk KJ. Upper limit ventricular stimulation in respiratory rate responsive pacing due to electrocautery. Pacing Clin Electrophysiol. ;12:1723-1720.
Chauvin M, Crenner F, Brechenmacher C. Interaction between permanent cardiac pacing and electrocautery: the significance of electrode position. Pacing Clin Electrophysiol. ;15:2033-2028.
Levine PA, Balady GJ, Lazar HL. Electrocautery and pacemakers: management of the paced patient subject to electrocautery. Ann Thorac Surg. ;41:313-317.
Erdman S, Levinsky L, Servadio C. Safety precautions in the management of patients with pacemakers when electrocautery operations are performed. Surg Gynecol Obstet. ;167:311-314.
Alexopoulos GS, Frances RJ. ECT and cardiac patients with pacemakers. Am J Psychiatry. ;137:1111-1112.
Rozner MA. Intrathoracic gadgets: update on pacemakers and implantable cardioverter-defibrillators. :212.
Salukhe TV, Dob D, Sutton R. Pacemakers and defibrillators: anaesthetic implications. Br J Anaesth. ;93:95-104.
Philbin DM, Marieb MA, Aithal KH. Inappropriate shocks delivered by an ICD as a result of sensed potentials from a transcutaneous electronic nerve stimulation unit. Pacing Clin Electrophysiol. ;21:2011-2010.
Raitt MH, Stelzer KJ, Laramore GE. Runaway pacemaker during high-energy neutron radiation therapy. Chest. ;106:955-957.
Souliman SK, Christie J. Pacemaker failure induced by radiotherapy. Pacing Clin Electrophysiol. ;17:270-273.
Muller- Runkel R, Orsolini G, Kalokhe UP. Monitoring the radiation dose to a multiprogrammable pacemaker during radical radiation therapy: a case report. Pacing Clin Electrophysiol. ;13:1470-1466.
Langberg J, Abber J, Thuroff JW. The effects of extracorporeal shock wave lithotripsy on pacemaker function. Pacing Clin Electrophysiol. ;10:1142-1146.
Cooper D, Wilkoff B, Masterson M. Effects of extracorporeal shock wave lithotripsy on cardiac pacemakers and its safety in patients with implanted cardiac pacemakers. Pacing Clin Electrophysiol. ;11:1386-1387.
Albers DD, Lybrand FE, Axton JC. Shockwave lithotripsy and pacemakers: experience with 20 cases. J Endourol. ;9:301-303.
Ganem JP, Carson CC. Cardiac arrhythmias with external fixed rate signal generators in shock wave lithotripsy with the Medstone lithotripter. Urology. ;51:548-552.
Ito S, Shibata H, Okahisa T. Endoscopic therapy using monopolar and bipolar snare with a high-frequency current in patients with a pacemaker. Endoscopy. ;26:270.
Aylward P, Blood R, Tonkin A. Complications of defibrillation with permanent pacemaker in situ. Pacing Clin Electrophysiol. ;2:464-462.
Waller C, Callies F, Langenfeld H. Adverse effects of direct current cardioversion on cardiac pacemakers and electrodes: is external cardioversion contraindicated in patients with permanent pacing systems?. Europace. ;6:165-168.
Das G, Eaton J. Pacemaker malfunction following transthoracic countershock. Pacing Clin Electrophysiol. ;4:487-490.
Levine PA, Barold SS, Fletcher RD. Adverse acute and chronic effects of electrical defibrillation and cardioversion on implanted unipolar cardiac pacing systems. J Am Coll Cardiol. ;1:1422-1413.
Gould L, Patel S, Gomes GI. Pacemaker failure following external defibrillation. Pacing Clin Electrophysiol. ;4:577-575.
Calkins H, Brinker J, Veltri EP. Clinical interactions between pacemaker and automated implantable cardioverter defibrillators. J Am Coll Cardiol. ;16:666-673.
Yee R, Jones DL, Klein GJ. Pacing threshold changes after transvenous catheter countershock. Am J Cardiol. ;53:507-503.
Lau FY, Bilitch M, Wintroub HJ. Protection of implanted pacemakers from excessive electrical energy of DC shock. Am J Cardiol. ;23:244-249.
Mehta Y, Swaminathan M, Juneja R. Noncardiac surgery and pacemaker cardioverter-defibrillator management. J Cardiothorac Vasc Anesth. ;12:224-221.
Mangar D, Atlas GM, Kane PB. Electrocautery-induced pacemaker malfunction during surgery. Can J Anaesth. ;38:618-616.
Heymsfield SB, Wang Z, Visser M. Techniques used in the measurement of body composition: an overview with emphasis on bioelectrical impedance analysis. Am J Clin Nutr. ;64:478S-484S.
Leeab SY, Gallagherc D. Assessment methods in human body composition. Curr Opin Clin Nutr Metab Care. ;11:572-566.
Rienzo VD, Minelli M, Sambugaro R. Applicability of extra- cellular electrical impedance tomography in monitoring respiratory tract inflammation. J Investig Allergol Clin Immunol. ;17:38-34.
Dohrmann ML, Goldschlager NF. Myocardial stimulation threshold in patients with cardiac pacemakers: effect of physiologic variables, pharmacologic agents, and lead electrodes. Cardiol Clin. ;3:537-527.
Saxon LA, Stevenson WG, Middlekauff HR. Increased risk of progressive hemodynamic deterioration in advanced heart failure patients requiring permanent pacemakers. Am Heart J. ;125:1306-1310.
Fleisher LA, Beckman JA, Brown KA. ACC/AHA 2007 guidelines on perioperative cardiovascular evaluation and care for noncardiac surgery. .
Simon AB. Perioperative management of the pacemaker patient. Anesthesiology. 1977;46:127-31; Simon AB. Perioperative management of the pacemaker patient. Circulation. ;116:1976-1971.
Zaidan JR. Pacemakers. Anesthesiology. ;60:334-319.
Kemnitz J, Peters J. Cardiac pacemakers and implantable cardioverter defibrillators in the perioperative phase. Anasthesiol Intensivmed Notfallmed Schmerzther. ;28:212-199.
Finfer SR. Pacemaker failure on induction of anaesthesia. Br J Anaesth. ;66:512-509.
Atlee JL, Bernstein AD. Cardiac rhythm management devices, perioperative management. Anesthesiology. ;95:1492-1506.
Lamas GA, Rebecca GS, Braunwald NS. Pacemaker malfunction after nitrous oxide anesthesia. Am J Cardiol. ;56:995-996.
Thiagarajah S, Azar I, Agres M. Pacemaker malfunction associated with positive pressure ventilation. Anesthesiology. ;58:565-566.